Selecting Optimal Proton Pencil Beam Scanning Plan Parameters to Reduce Dose Discrepancy between Discrete Spot Plan and Continuous Scanning: A Proof-of-Concept Study


ABSTRACT:

Simple Summary

Proton beam therapy delivered via pencil beam scanning is an advanced method for treating cancer. Two types of scanning modes are available for this method: discrete spot scanning (DSS) and continuous scanning. Continuous scanning has some advantages over DSS including improved proton beam delivery efficiency and reduced delivery time. However, continuous scanning delivers a dose between consecutive spots, and currently commercially available treatment planning systems do not account for this. Consequently, the continuous scanning delivered dose is inherently different from the planned one, which is discrete spot based. We conducted a proof-of-concept study of the ability of a prediction model to reduce the dose discrepancy between continuous scanning and discrete s