Ontology highlight
ABSTRACT: Background
Chest compression and defibrillation are essential components of cardiac arrest treatment. Mechanical chest compression devices (MCCD) and automated external defibrillators (AED) are used separately in clinical practice. We developed an automated compression-defibrillation apparatus (ACDA) that performs mechanical chest compression and automated defibrillation. We investigated the performance of cardiopulmonary resuscitation (CPR) with automatic CPR (A-CPR) compared to that with MCCD and AED (conventional CPR: C-CPR).Methods
Pigs were randomized into A-CPR or C-CPR groups: The A-CPR group received CPR+ACDA, and the C-CPR group received CPR+MCCD+AED. Hemodynamic parameters, outcomes, and time variables were measured. During a simulation study, healthcare providers
SUBMITTER: Jung WJ
PROVIDER: S-EPMC10455516 | biostudies-literature | 2023 Aug
REPOSITORIES: biostudies-literature