A valve embolized twice: a case report of Sapien 3 embolization in left ventricle and left atrium during transcatheter transapical mitral valve-in-valve implantations for a failed bioprosthesis.


ABSTRACT:

Background

Transcatheter mitral valve-in-valve (TMVIV) using the Sapien 3 (Edwards Lifesciences, Irvine, CA, USA) transcatheter heart valve is associated with high technical success and sustained valve performance. However, complications may occur or be detected during or after the procedure.

Case summary

We herein describe a rare case of a 59-year-old female who underwent TMVIV for a failed surgical mitral bioprosthesis. During the procedure, the device was embolized twice into the left ventricle and left atrium, respectively, resulting from the crimped transcatheter valve partially detaching from the balloon of the Certitude delivery system during passage through a tight transapical sheath. Fortunately, we were able to catch the dislodged valve and anchor it by the partial