A simulated comparison of lung tumor target verification using stereoscopic tomosynthesis or radiography.


ABSTRACT:

Purpose

Mobile lung tumors are increasingly being treated with ablative radiotherapy, for which precise motion management is essential. In-room stereoscopic radiography systems are able to guide ablative radiotherapy for stationary cranial lesions but not optimally for lung tumors unless fiducial markers are inserted. We propose augmenting stereoscopic radiographic systems with multiple small x-ray sources to provide the capability of imaging with stereoscopic, single frame tomosynthesis.

Methods

In single frame tomosynthesis, nine x-ray sources are placed in a 3 × 3 configuration and energized simultaneously. The beams from these sources are collimated so that they converge on the tumor and then diverge to illuminate nine non-overlapping sectors on the detector. These nine