ABSTRACT: Background
In transcatheter aortic valve implantation (TAVI) using a SAPIEN3 balloon-expandable valve (S3), wire withdrawal from the left ventricle (LV) during the procedure before deployment can induce vascular injury in the access site or require surgical treatment when an S3 removal is attempted. We present a successful case of bailout from this situation safely with a minimally invasive technique using a 6-F snare catheter (SC). Case summary
An 86-year-old woman with severe aortic stenosis underwent trans-femoral TAVI using an S3 under conscious sedation. After a pre-shaped wire was inserted into the LV from the right femoral artery, the LV wire was accidentally withdrawn completely from the LV before deployment. Wire re-insertion using a soft straight wire t