Intra-aortic and Intra-caval Balloon Pump Devices in Experimental Non-traumatic Cardiac Arrest and Cardiopulmonary Resuscitation.


ABSTRACT: Intra-aortic balloon pump (IABP) use during CPR has been scarcely studied. Intra-caval balloon pump (ICBP) may decrease backward venous flow during CPR. Mechanical chest compressions (MCC) were initiated after 10 min of cardiac arrest in anesthetized pigs. After 5 min of MCC, IABP (n = 6) or ICBP (n = 6) was initiated. The MCC device and the IABP/ICBP had slightly different frequencies, inducing a progressive peak pressure phase shift. IABP inflation 0.15 s before MCC significantly increased mean arterial pressure (MAP) and carotid blood flow (CBF) compared to inflation 0.10 s after MCC and to MCC only. Coronary perfusion pressure significantly increased with IABP inflation 0.25 s before MCC compared to inflation at MCC. ICBP inflation before MCC significantly increased MAP and CBF compare