ABSTRACT: Background
Compliance with dialysis fluid ultrapurity standards is a paramount for online modalities. More than 200 dialysis fluid samples have been analyzed monthly for years in our two dialysis units, with compliant microbiological results until mid-2020.Aim
In mid-2020, an unusual occurrence (30%) of contaminated dialysis fluids in dialysis units led us to investigate to determine the source.Methods
Microbiological methods for aquaphilic bacteria culturing and endotoxin detection in dialysis fluids were routinely performed on a monthly basis for all dialysis machines. As the contamination appeared randomly and almost simultaneously in our two units without any routine change or febrile syndrome, we searched for a common cause. Supplier's sampling kits as well as