Evaluating the Efficiency of the Cobas 6800 System for BK Virus Detection in Plasma and Urine Samples.
Ontology highlight
ABSTRACT: We evaluated the overall performance of the Cobas 6800 BKV test in detecting BK virus (BKV). We examined the imprecision of the Cobas 6800 BKV test and compared the qualitative and quantitative results obtained from the Cobas 6800 BKV test and the Real-Q BKV quantification assay. We assessed 88 plasma and 26 urine samples collected between September and November 2022 from patients with BKV infection using the Real-Q BKV quantitative assay. The lognormal coefficient of variation indicated that the inter-assay precision of the Cobas 6800 BKV test ranged from 13.86 to 33.83%. A strong correlation was observed between the quantitative results obtained using the Cobas 6800 BKV test and the Real-Q BKV quantification assay for plasma samples. The Spearman's rank correlation coefficients (ρ) for p