Unknown

Dataset Information

Digital process control of multi-step assays on centrifugal platforms using high-low-high rotational-pulse triggered valving.


ABSTRACT: Due to their capability for comprehensive sample-to-answer automation, the interest in centrifugal microfluidic systems has greatly increased in industry and academia over the last quarter century. The main applications of these "Lab-on-a-Disc" (LoaD) platforms are in decentralised bioanalytical point-of-use / point-of-care testing. Due to the unidirectional and omnipresent nature of the centrifugal force, advanced flow control is key to coordinate multi-step / multi-reagent assay formats on the LoaD. Formerly, flow control was often achieved by capillary burst valves which require gradual increments of the spin speed of the system-innate spindle motor. Recent advanced introduced a flow control scheme called 'rotational pulse actuated valves'. In these valves the sequence of valve actuatio

SUBMITTER: Early PL 

PROVIDER: S-EPMC10490917 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature