Discrepancy between two invasive blood pressure measurements in patients receiving intra-aortic balloon pump therapy.


ABSTRACT:

Background

Hemodynamic monitoring is imperative for patients with cardiogenic shock undergoing Intra-aortic Balloon Pump (IABP) therapy. Blood pressure monitoring encompasses non-invasive, invasive peripheral arterial pressure (IPAP), and invasive central aortic pressure (ICAP) methods. However, marked disparities exist between IPAP and ICAP. This study examined the discrepancies between IPAP and ICAP and their clinical significance.

Methods

A retrospective analysis was conducted on cardiogenic shock patients who underwent IABP therapy and were admitted to the Coronary Care Unit (CCU) of a tertiary hospital in China from March 2017 to November 2022. The Bland-Altman plot illustrated the discrepancy between IPAP and ICAP. A clinically significant difference between ICAP and I