Efficacy of Paired Electrochemical Sensors for Measuring Ozone Concentrations.
Ontology highlight
ABSTRACT: Typical low-cost electrochemical sensors for ozone (O3) are also highly responsive to nitrogen dioxide (NO2). Consequently, a single sensor's response to O3 is indistinguishable from its response to NO2. Recently, a method for quantifying O3 concentrations became commercially available (Alphasense Ltd., Essex, UK): collocating a pair of sensors, a typical oxidative gas sensor that responds to both O3 and NO2 (model OX-B431) and a second similar sensor that filters O3 and responds only to NO2 (model NO2-B43F). By pairing the two sensors, O3 concentrations can be calculated. We calibrated samples of three NO2-B43F sensors and three OX-B431 sensors with NO2 and O3
SUBMITTER: Zuidema C
PROVIDER: S-EPMC10494896 | biostudies-literature | 2019 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA