ABSTRACT: Background
Broad generalization of radiomics-assisted models may be impeded by concerns about variability. This study aimed to evaluate the merit of combatting batch effect (ComBat) harmonization in reducing the variability of voxel size-related radiomics in both phantom and clinical study in comparison with image resampling correction method.Methods
A pulmonary phantom with 22 different types of nodules was scanned by computed tomography (CT) with different voxel sizes. The variability of voxel size-related radiomics features was evaluated using concordance correlation coefficient (CCC), dynamic range (DR), and intraclass correlation coefficient (ICC). ComBat and image resampling compensation methods were used to reduce variability of voxel size-related radiomics. The perc