Geometric accuracy in patient positioning for stereotactic radiotherapy of intracranial tumors.


ABSTRACT:

Background/purpose

This study determines and compares the geometric setup errors between stereoscopic x-ray and kilo-voltage cone beam CT (CBCT) in phantom tests on a linear accelerator (linac) for image-guided (IG) stereotactic radiotherapy of intracranial tumors. Additionally, dose-volume metrics in the target volumes of the setup errors of CBCT were evaluated.

Materials/methods

A Winston-Lutz- and an anthropomorphic phantom were used. The mean deviation and root mean square error (RMSE) of CBCT and stereoscopic x-ray were compared. Dose-volume metrics of the planning target volume (PTV) and gross target volume (GTV) for CBCT were calculated.

Results

The RMSEs in the tests with the Winston-Lutz-Phantom were 0.3 mm, 1.1 mm and 0.3 mm for CBCT and 0.1 mm, 0,1 mm and <