Dynamic Cyclic Fatigue and Differential Scanning Calorimetry Analysis of R-Motion.


ABSTRACT:

Introduction

The aim of this study was to assess the dynamic cyclic fatigue resistance of an R-Motion file at simulated body temperature and document corresponding phase transformations compared to OneCurve and HyFlex EDM (HFEDM).

Methods

R-Motion (25/.06), OneCurve (25/.06), and HFEDM (25/.06) files were selected and divided into 3 groups (n = 9) according to the file type. Dynamic cyclic fatigue testing was done with a custom-made artificial stainless-steel canal that had a 90° angle of curvature and a 5-mm radius of curvature. Files were operated continuously at body temperature until fracture in the artificial canal. The time to fracture was calculated. Statistical analysis was performed, and significance was set at 5%. Phase transformation temperatures for 2 instruments