Comparative study of sub-second temporal resolution 4D-MRI and 4D-CT for target motion assessment in a phantom model.
Ontology highlight
ABSTRACT: To develop and investigate the feasibility of sub-second temporal resolution volumetric T1-weighted four-dimensional (4D-) MRI in comparison with 4D-CT for respiratory-correlated motion assessment using an MRI/CT-compatible phantom. Sub-second high temporal resolution (0.5 s) gradient-echo T1-weighted 4D-MRI was developed using a volumetric acquisition scheme with compressed sensing. An MRI/CT-compatible motion phantom (simulated liver tumor) with three sinusoidal movements of amplitudes and two respiratory patterns was introduced and imaged with 4D-MRI and 4D-CT to investigate the geometric accuracy of the target movement. The geometric accuracy, including centroid position, volume, similarity index of dice similarity coefficient (DSC), and Hausdorff distance (HD), was systematically eval