Agreement and precision of wide and cube scan measurements between swept-source and spectral-domain OCT in normal and glaucoma eyes.


ABSTRACT: This study aimed to evaluate agreement of Wide scan measurements from swept-source optical coherence tomography (SS-OCT) Triton and spectral-domain OCT (SD-OCT) Maestro in normal/glaucoma eyes, and to assess the precision of measurements from Wide and Cube scans of both devices. Three Triton and three Maestro operator/device configurations were created by pairing three operators, with study eye and testing order randomized. Three scans were captured for Wide (12 mm × 9 mm), Macular Cube (7 mm × 7 mm-Triton; 6 mm × 6 mm-Maestro), and Optic Disc Cube (6 mm × 6 mm) scans for 25 normal eyes and 25 glaucoma eyes. Parameter measurements included circumpapillary retinal nerve fiber layer(cpRNFL), ganglion cell layer + inner plexiform layer (GCL+), and ganglion cell complex (GCL++). A two-way random effect analysis of variance model was used to estimate the repeatability and reproducibility; agreement was evaluated by Bland-Altman analysis and Deming regression. The precision estimates were low, indicating high precision, for all thickness measurements with the majority of the limits < 5 µm for the macula and < 10 µm for the optic disc. Precision of the Wide and Cube scans were comparable. Excellent agreement between the two devices was found for Wide scans, with the mean difference < 3 µm across all measurements (cpRNFL < 3 µm, GCL+  < 2 µm, GCL ++  < 1 µm), indicating interoperability. A single Wide scan covering the peripapillary and macular regions may be useful for glaucoma diagnosis and management.

SUBMITTER: Hou H 

PROVIDER: S-EPMC10517954 | biostudies-literature | 2023 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Agreement and precision of wide and cube scan measurements between swept-source and spectral-domain OCT in normal and glaucoma eyes.

Hou Huiyuan H   El-Nimri Nevin W NW   Durbin Mary K MK   Arias Juan D JD   Moghimi Sasan S   Weinreb Robert N RN  

Scientific reports 20230923 1


This study aimed to evaluate agreement of Wide scan measurements from swept-source optical coherence tomography (SS-OCT) Triton and spectral-domain OCT (SD-OCT) Maestro in normal/glaucoma eyes, and to assess the precision of measurements from Wide and Cube scans of both devices. Three Triton and three Maestro operator/device configurations were created by pairing three operators, with study eye and testing order randomized. Three scans were captured for Wide (12 mm × 9 mm), Macular Cube (7 mm ×  ...[more]

Similar Datasets

| S-EPMC6705867 | biostudies-literature
| S-EPMC6137103 | biostudies-literature
| S-EPMC3169218 | biostudies-other
| S-EPMC11487269 | biostudies-literature
| S-EPMC7442878 | biostudies-literature
| S-EPMC8589877 | biostudies-literature
| S-EPMC7285531 | biostudies-literature
| S-EPMC10757970 | biostudies-literature
| S-EPMC9405474 | biostudies-literature
| S-EPMC4433247 | biostudies-literature