ABSTRACT: Purpose
To identify risk factors of ocular anterior segment measurement error by the IOLMaster 700 in eyes implanted with an implantable Collamer lens (ICL).Methods
In total, 152 patients with clear lens (152 eyes, group 1) and another 32 cataract patients (57 eyes, group 2) who underwent ICL implantation were included, and the presence of measurement error by the IOLMaster 700 was determined based on B-scan images. The risk factors for measurement error were evaluated by logistic regression, and the optimal threshold was determined using receiver operating characteristic analysis.Results
The ICL was misidentified as the anterior surface of the crystalline lens in 51.97% of eyes (79/152) in group 1 and 80.70% of eyes (46/57) in group 2. For every 100-µm decrease in the vault height, a 3.57- and 5.78-fold increase in the risk of measurement error was observed in group 1 and group 2, respectively. We identified an optimal threshold of the vault height at 389.47 µm for predicting biometric measurement error in eyes implanted with ICL, which showed an area under the curve of 0.93 (95% confidence interval, 0.90-0.97), a sensitivity of 0.87, and a specificity of 0.86.Conclusions
Patients with ICL implantation, particularly those with a vault height less than 389.47 µm, are at a greater risk of anterior segment biometric measurement error by the IOLMaster 700.Translational relevance
The threshold of vault height can help to identify high-risk patients and further optimize biometric measurement.
SUBMITTER: Zhang J
PROVIDER: S-EPMC10519433 | biostudies-literature | 2023 Sep
REPOSITORIES: biostudies-literature

Zhang Jiaqing J Chen Fei F Han Xiaotong X Qiu Xiaozhang X Liu Zhenzhen Z Chen Xiaoyun X Jin Guangming G Qu Bo B Yao Huan H Ye Yiming Y Yu Keming K Tan Xuhua X Luo Lixia L
Translational vision science & technology 20230901 9
<h4>Purpose</h4>To identify risk factors of ocular anterior segment measurement error by the IOLMaster 700 in eyes implanted with an implantable Collamer lens (ICL).<h4>Methods</h4>In total, 152 patients with clear lens (152 eyes, group 1) and another 32 cataract patients (57 eyes, group 2) who underwent ICL implantation were included, and the presence of measurement error by the IOLMaster 700 was determined based on B-scan images. The risk factors for measurement error were evaluated by logisti ...[more]