ABSTRACT: Background
Kinetic modelling of dynamic PET typically requires knowledge of the arterial radiotracer concentration (arterial input function, AIF). Its accurate determination is very difficult in mice. AIF measurements in an extracorporeal shunt can be performed; however, this introduces catheter dispersion. We propose a framework for extracorporeal dispersion correction and validated it by comparison to invasively determined intracorporeal AIFs using implanted microprobes.Results
The response of an extracorporeal radiation detector to radioactivity boxcar functions, characterised by a convolution-based dispersion model, gave best fits using double-gamma variate and single-gamma variate kernels compared to mono-exponential kernels for the investigated range of flow rates. Pa