Clinical Performance of the Gore Septal Occluder in Patent Foramen Ovale Closure in Different Septal Anatomies: 1-Year Results from a Single-Center Experience.


ABSTRACT:

Background

PFO (Patent foramen ovale) is a common defect that affects about 25% of the population. Although its presence is asymptomatic in the majority of the cases, the remaining part becomes overt with different symptoms, including cryptogenic stroke. PFO closure is currently a widely available procedure in complex anatomy, with Amplatzer PFO Occluder (APO) being the most commonly used tool. However, the performance of another device, the GORE Septal Occluder (GSO), has not been completely explored with regard to different septal anatomies.

Methods

From March 2012 to June 2020, 118 consecutive patients with an indication of PFO closure were treated using the GSO system, included in a prospective analysis, and followed. After 12 months, every patient underwent transcranial Doppler ultrasound to evaluate the effectiveness of treatment.

Results

Of 111 patients evaluated, 107 showed effective PFO closure (96.4%), and 4 showed a residual shunt (3.6%). To better evaluate the device performance, the overall population was sorted into two clusters based on the echocardiographic characteristics. The main difference between groups was for PFO width (4.85 ± 1.8 vs. 2.9 ± 1 mm, p < 0.001) and PFO tunnel length (12.6 ± 3.8 vs. 7.2 ± 2, p < 0.001), allowing complex and simple anatomies to be identified, respectively. Regardless of the aforementioned cluster, the GSO performance required to reach an effective closure was independent of anatomy type and the chosen device size.

Conclusion

The GSO device showed a high closure rate at 1-year follow-up in patients, with at least one anatomical factor of complexity of PFO irrespective of the level of complexity itself.

SUBMITTER: Verolino G 

PROVIDER: S-EPMC10531839 | biostudies-literature | 2023 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Clinical Performance of the Gore Septal Occluder in Patent Foramen Ovale Closure in Different Septal Anatomies: 1-Year Results from a Single-Center Experience.

Verolino Giuseppe G   Calderone Dario D   Gavazzoni Mara M   Sala Davide D   Sganzerla Paolo P  

Journal of clinical medicine 20230913 18


<h4>Background</h4>PFO (Patent foramen ovale) is a common defect that affects about 25% of the population. Although its presence is asymptomatic in the majority of the cases, the remaining part becomes overt with different symptoms, including cryptogenic stroke. PFO closure is currently a widely available procedure in complex anatomy, with Amplatzer PFO Occluder (APO) being the most commonly used tool. However, the performance of another device, the GORE Septal Occluder (GSO), has not been compl  ...[more]