Development and Analytical Evaluation of a Point-of-Care Electrochemical Biosensor for Rapid and Accurate SARS-CoV-2 Detection.


ABSTRACT: The COVID-19 pandemic has underscored the critical need for rapid and accurate screening and diagnostic methods for potential respiratory viruses. Existing COVID-19 diagnostic approaches face limitations either in terms of turnaround time or accuracy. In this study, we present an electrochemical biosensor that offers nearly instantaneous and precise SARS-CoV-2 detection, suitable for point-of-care and environmental monitoring applications. The biosensor employs a stapled hACE-2 N-terminal alpha helix peptide to functionalize an in situ grown polypyrrole conductive polymer on a nitrocellulose membrane backbone through a chemical process. We assessed the biosensor's analytical performance using heat-inactivated omicron and delta variants of the SARS-CoV-2 virus in artificial saliva (AS) and nasal swab (NS) samples diluted in a strong ionic solution, as well as clinical specimens with known Ct values. Virus identification was achieved through electrochemical impedance spectroscopy (EIS) and frequency analyses. The assay demonstrated a limit of detection (LoD) of 40 TCID50/mL, with 95% sensitivity and 100% specificity. Notably, the biosensor exhibited no cross-reactivity when tested against the influenza virus. The entire testing process using the biosensor takes less than a minute. In summary, our biosensor exhibits promising potential in the battle against pandemic respiratory viruses, offering a platform for the development of rapid, compact, portable, and point-of-care devices capable of multiplexing various viruses. The biosensor has the capacity to significantly bolster our readiness and response to future viral outbreaks.

SUBMITTER: Meshesha M 

PROVIDER: S-EPMC10534802 | biostudies-literature | 2023 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Development and Analytical Evaluation of a Point-of-Care Electrochemical Biosensor for Rapid and Accurate SARS-CoV-2 Detection.

Meshesha Mesfin M   Sardar Anik A   Supekar Ruchi R   Bhattacharjee Lopamudra L   Chatterjee Soumyo S   Halder Nyancy N   Mohanta Kallol K   Bhattacharyya Tarun Kanti TK   Pal Biplab B  

Sensors (Basel, Switzerland) 20230920 18


The COVID-19 pandemic has underscored the critical need for rapid and accurate screening and diagnostic methods for potential respiratory viruses. Existing COVID-19 diagnostic approaches face limitations either in terms of turnaround time or accuracy. In this study, we present an electrochemical biosensor that offers nearly instantaneous and precise SARS-CoV-2 detection, suitable for point-of-care and environmental monitoring applications. The biosensor employs a stapled hACE-2 N-terminal alpha  ...[more]

Similar Datasets

| S-EPMC8106877 | biostudies-literature
| S-EPMC9352833 | biostudies-literature
| S-EPMC7586458 | biostudies-literature
| S-EPMC8447555 | biostudies-literature
| S-EPMC7864991 | biostudies-literature
| S-EPMC7827295 | biostudies-literature
| S-EPMC9197783 | biostudies-literature
| S-EPMC8405239 | biostudies-literature
| S-EPMC10296353 | biostudies-literature
| S-EPMC9880620 | biostudies-literature