Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats.


ABSTRACT: Previously, radioimmunoassay (RIA) has been the only assay to measure insulin-like growth factor-1 (IGF-1) to diagnose hypersomatotropism (HS). Due to radiation concerns, availability, and the cost of IGF-1 RIA, validation of assays for automated analysers such as a chemiluminescent immunoassay (CLIA) is needed. The aim of this study was to validate a CLIA for measurement of feline IGF-1 (IMMULITE 2000® XPi, Siemens Medical Solutions Diagnostics, Malvern, PA, USA) compared to IGF1 RIA, establish reference interval (RI), and determine a cut-off value for diagnosis of HS in diabetic cats. Validation of assay performance included precision, linearity, and recovery studies. Right-sided RI was determined using surplus serum of 50 healthy adult cats. Surplus serum samples of diabetic