On the Limit of Detection in Infrared Spectroscopic Imaging.


ABSTRACT: Infrared (IR) spectroscopic imaging instruments' performance can be characterized and optimized by an analysis of their limit of detection (LOD). Here we report a systematic analysis of the LOD for Fourier transform IR (FT-IR) and discrete frequency IR (DFIR) imaging spectrometers. In addition to traditional measurements of sample and blank data, we propose a decision theory perspective to pose the determination of LOD as a binary classification problem under different assumptions of noise uniformity and correlation. We also examine three spectral analysis approaches, namely, absorbance at a single frequency, average of absorbance over selected frequencies and total spectral distance - to suit instruments that acquire discrete or contiguous spectral bandwidths. The analysis is validated by

SUBMITTER: Lux L 

PROVIDER: S-EPMC10539114 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature