EDOF intraocular lens design: shift in image plane vs object vergence.


ABSTRACT:

Background

To compare 2 different design scenarios of EDOF-IOLs inserted in the Liou-Brennan schematic model eye using raytracing simulation as a function of pupil size.

Methods

Two EDOF IOL designs were created and optimized for the Liou-Brennan schematic model eye using Zemax ray tracing software. Each lens was optimized to achieve a maximum Strehl ratio for intermediate and far vision. In the first scenario, the object was located at infinity (O1), and the image plane was positioned at far focus (I1) and intermediate focus (I2) to emulate far and intermediate distance vision, respectively. In the second scenario, the image plane was fixed at I1 according to the first scenario. The object plane was set to infinity (O1) for far-distance vision and then shifted closer to the