Dynamic torque and screw-in force of four different glide path instruments assessed in simulated single- and double-curved canals: An in vitro study.


ABSTRACT:

Background/purpose

The glide path instruments are the introductory instruments into the canals; thus, they should be durable enough to withstand torsional stress/screw-in force. This study aimed to investigate the torque and screw-in force of TruNatomy Glider (TN), ProGLIDER (PG), Hyflex EDM (EDM) and Dent Craft RE (RE) glide path instruments in single- and double-curved canals.

Materials and methods

Each instrument brand was divided into two groups (n = 7 each) according to the canal configuration. Torque and screw-in force were recorded during automated instrumentation of simulated resin canals with XSmart IQ cordless motor after the canal patency was checked with a #10 K-file. The values were statistically analyzed using the Kolmogorov-Smirnov test followed by the Kruskal

SUBMITTER: Thu M 

PROVIDER: S-EPMC10548005 | biostudies-literature | 2023 Oct

REPOSITORIES: biostudies-literature