Method development for thermal desorption-gas chromatography-tandem mass spectrometry (TD-GC-MS/MS) analysis of trace level fluorotelomer alcohols emitted from consumer products.
Ontology highlight
ABSTRACT: The scientific foundation for per- and polyfluoroalkyl substances (PFAS) measurements in water, soils, sediments, biosolids, biota, and outdoor air has rapidly expanded; however, there are limited efforts devoted to developing analytical methods to measure vapor-phase PFAS in indoor air. A gas chromatography-tandem mass spectrometry (GC-MS/MS) method coupled with thermal desorption (TD) sorbent tube analysis was developed to quantify trace levels of fluorotelomer alcohols (FTOHs) emitted from consumer products in the indoor environment. Method evaluation included determination of instrument detection limits (IDLs), quality assurance checks of target standards purchased from different vendors, sample loss during storage, and TD sorbent breakthrough with tubes coupled in-series. The IDLs for