Safety and function of programmable ventriculo-peritoneal shunt valves: An in vitro 7 Tesla magnetic resonance imaging study.


ABSTRACT:

Objective

The quantity of ultra-high field MRI neuroimaging studies has rapidly increased. This study tests function, safety, and image artifacts of two frequently implanted programmable ventriculo-peritoneal (VP) shunt valves in a 7T MRI system.

Methods

All tests were performed using a whole-body 7T MRI system. Three proGAV 2.0 and 3 CODMAN CERTAS® Plus programmable VP-shunt valves were tested in three steps. 1) Deflection angle tests close to the bore opening at the location of a static magnetic field gradient of 3-5 T/m. 2) Valves were fixed on a spherical phantom in 3 positions (a. lateral, b. cranial, c. cranial with 22.5° tilt anteriorly) and assessed for keeping the programmed pressure setting and reprogrammability. 3) Valves were fixed on the phantom and positioned l