Single pulse protoacoustic range verification using a clinical synchrocyclotron.
Ontology highlight
ABSTRACT: Objective.Proton therapy as the next generation radiation-based cancer therapy offers dominant advantages over conventional radiation therapy due to the utilization of the Bragg peak; however, range uncertainty in beam delivery substantially mitigates the advantages of proton therapy. This work reports using protoacoustic measurements to determine the location of proton Bragg peak deposition within a water phantom in real time during beam delivery.Approach.In protoacoustics, proton beams have a definitive range, depositing a majority of the dose at the Bragg peak; this dose is then converted to heat. The resulting thermoelastic expansion generates a 3D acoustic wave, which can be detected by acoustic detectors to localize the Bragg peak.Main results.Protoacoustic measu
SUBMITTER: Caron J
PROVIDER: S-EPMC10567060 | biostudies-literature | 2023 Feb
REPOSITORIES: biostudies-literature