Comparison of diffusion weighted imaging b0 with T2*-weighted gradient echo or susceptibility weighted imaging for intracranial hemorrhage detection after reperfusion therapy for ischemic stroke.


ABSTRACT:

Purpose

Diffusion-weighted imaging (DWI) b0 may be able to substitute T2*-weighted gradient echo (GRE) or susceptibility-weighted imaging (SWI) in case of comparable detection of intracranial hemorrhage (ICH), thereby reducing MRI examination time. We evaluated the diagnostic accuracy of DWI b0 compared to T2*GRE or SWI for detection of ICH after reperfusion therapy for ischemic stroke.

Methods

We pooled 300 follow-up MRI scans acquired within 1 week after reperfusion therapy. Six neuroradiologists each rated DWI images (b0 and b1000; b0 as index test) of 100 patients and, after a minimum of 4 weeks, T2*GRE or SWI images (reference standard) paired with DWI images of the same patients. Readers assessed the presence of ICH (yes/no) and type of ICH according to the Heidelberg