In Vitro Validation of a Novel Continuous Intra-Abdominal Pressure Measurement System (TraumaGuard).
Ontology highlight
ABSTRACT: Introduction: Intra-abdominal pressure (IAP) has been recognized as an important vital sign in critically ill patients. Due to the high prevalence and incidence of intra-abdominal hypertension in surgical (trauma, burns, cardiac) and medical (sepsis, liver cirrhosis, acute kidney injury) patients, continuous IAP (CIAP) monitoring has been proposed. This research was aimed at validating a new CIAP monitoring device, the TraumaGuard from Sentinel Medical Technologies, against the gold standard (height of a water column) in an in vitro setting and performing a comparative analysis among different CIAP measurement technologies (including two intra-gastric and two intra-bladder measurement devices). A technical and clinical guideline addressing the strengths and weaknesses of each device
SUBMITTER: Tayebi S
PROVIDER: S-EPMC10573363 | biostudies-literature | 2023 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA