Unknown

Dataset Information

MiniMed 780G™ advanced hybrid closed-loop system performance in Egyptian patients with type 1 diabetes across different age groups: evidence from real-world users.


ABSTRACT:

Background

Advanced hybrid closed loop (AHCL) system provides both automated basal rate and correction boluses to keep glycemic values in a target range.

Objectives

To evaluate the real-world performance of the MiniMed™ 780G system among different age groups of Egyptian patients with type 1diabetes.

Methods

One-hundred seven AHCL system users aged from 3 to 71 years were enrolled. Data uploaded by patients were aggregated and analyzed. The mean glucose management indicator (GMI), percentage of time spent within glycemic ranges (TIR), time below range (TBR) and time above range (TAR) were determined.

Results

Six months after initiating Auto Mode, patients spent a mean of 85.31 ± 22.04% of the time in Auto Mode (SmartGuard) and achieved a mean GMI of 6.95 ± 0.58%

SUBMITTER: Elbarbary NS 

PROVIDER: S-EPMC10580510 | biostudies-literature | 2023 Oct

REPOSITORIES: biostudies-literature