Effect and Safety of Pressure Sensor-equipped Handpiece in Phacoemulsification System.


ABSTRACT:

Purpose

To evaluate the effect of the Active Sentry handpiece of the Centurion Vision System compared to the Centurion Ozil handpiece for phacoemulsification in cataract surgery.

Methods

A retrospective study was conducted on 281 patients (449 eyes) who underwent cataract surgery between August 2020 and June 2021. Preoperative measurements, intraoperative parameters, complication rate, and postoperative outcomes were compared between the Active Sentry handpiece and the Centurion Ozil handpiece groups. Additionally, the parameters were compared in different cataract severity groups and multiple predictive factors for the number of active surge mitigation (ASM) actuations were assessed with the Active Sentry handpiece.

Results

There were 198 eyes in the Active Sentry gr