Leadless epicardial pacing at the left ventricular apex: an animal study.


ABSTRACT:

Aims

State-of-the-art pacemaker implantation technique in infants and small children consists of pace/sense electrodes attached to the epicardium and a pulse generator in the abdominal wall with a significant rate of dysfunction during growth, mostly attributable to lead failure. In order to overcome lead-related problems, feasibility of epicardial implantation of a leadless pacemaker at the left ventricular apex in a growing animal model was studied.

Methods and results

Ten lambs (median body weight 26.8 kg) underwent epicardial implantation of a Micra transcatheter pacing system (TPS) pacemaker (Medtronic Inc., Minneapolis, USA). Using a subxyphoid access, the Micra was introduced through a short, thick-walled tube to increase tissue contact and to prevent tilting from the