Investigation of Platelet Function Analyzer 200 platelet function measurements in healthy cats and cats receiving clopidogrel.
Ontology highlight
ABSTRACT: The Platelet Function Analyzer 200 (PFA-200; Siemens) is an in vitro substitute for in vivo bleeding time that is designed to investigate platelet function in a more physiologic manner than traditional aggregometry. The analyzer reports a closure time (CT) as a marker of platelet function, and may also report the calculated platelet function measurement primary hemostasis components, PHC1 and PHC2. These incorporate the measured total volume (TV) of blood aspirated and the initial flow rate (IF). We determined, for the COL/ADP and P2Y cartridges, the median total volume (TVmedian), and RIs for CT, IF, TV, PHC1, and PHC2, and investigated the sensitivity and specificity of those parameters at the determined interpretation thresholds in determination of the clopidogrel effect. Hea