Design and Development of a Clot Burst Pressure Device to Investigate Resuscitation Strategies.


ABSTRACT:

Introduction

A reduction in clot strength is a hallmark feature of trauma-induced coagulopathy. A better understanding of clot integrity can optimize resuscitation strategies. We designed a device to gauge clot strength by pressurizing fluids over a formed clot and measuring the pressure needed to dislodge the clot. We hypothesized that this device could distinguish between clots formed in hypocoagulable and hypercoagulable states by observing differences in the clot burst pressure.

Methods

Whole blood from healthy volunteers was collected into sodium citrate tubes and was treated with heparin or fibrinogen to generate clots in a hypocoagulable or hypercoagulable state, respectively. Small bore holes were drilled into polystyrene plates, and recalcified blood was pipetted in