Establishing the best combination of the kappa free light chain index and oligoclonal bands for an accurate diagnosis of multiple sclerosis.


ABSTRACT:

Introduction

The immunoglobulin kappa free light chain (KFLC) index has been proposed as a potentially suitable alternative to oligoclonal IgG bands (OCGB) for diagnosing multiple sclerosis (MS), offering automation and reduced processing time. However, there is no consensus on the preferred approach or how to combine both techniques.

Methods

This prospective cohort study aimed to determine the best utilization of OCGB and KFLC index in patients with a clinically isolated syndrome (CIS) followed for at least two years. OCGB and KFLC were assessed using isoelectric focusing and immunoblotting and turbidimetry, respectively. Sensitivity, specificity, and accuracy for diagnosing MS were calculated for each method.

Results

The study included 371 patients, with 260 (70.1 %