Open-path measurement of stable water isotopologues using mid-infrared dual-comb spectroscopy.
Ontology highlight
ABSTRACT: We present an open-path mid-infrared dual-comb spectroscopy (DCS) system capable of precise measurement of the stable water isotopologues H216O and HD16O. This system ran in a remote configuration at a rural test site for 3.75 months with 60% uptime and achieved a precision of < 2‰ on the normalized ratio of H216O and HD16O (δD) in 1000s. Here, we compare the δD values from the DCS system to those from the National Ecological Observatory Network (NEON) isotopologue point sensor network. Over the multi-month campaign, the mean difference between the DCS δD values and the NEON δD values from a similar ecosystem is < 2‰ with a standard deviation of 18‰, which demonstrates the inherent accuracy of DCS measurements over a variety