Effect of metal implants and metal artifacts on back-projected two-dimensional entrance fluence determined by EPID dosimetry.


ABSTRACT:

Purpose

To evaluate the errors caused by metal implants and metal artifacts in the two-dimensional entrance fluences reconstructed using the back-projection algorithm based on electronic portal imaging device (EPID) images.

Methods

The EPID in the Varian VitalBeam accelerator was used to acquire portal dose images (PDIs), and then commercial EPID dosimetry software was employed to reconstruct the two-dimensional entrance fluences based on computed tomography (CT) images of the head phantoms containing interchangeable metal-free/titanium/aluminum round bars. The metal-induced errors in the two-dimensional entrance fluences were evaluated by comparing the γ results and the pixel value errors in the metal-affected regions. We obtained metal-artifact-free CT images by replacing

SUBMITTER: Cao Z 