Technical note: A collision prediction tool using Blender.
Ontology highlight
ABSTRACT: Non-coplanar radiotherapy treatment techniques on C-arm linear accelerators have the potential to reduce dose to organs-at-risk in comparison with coplanar treatment techniques. Accurately predicting possible collisions between gantry, table and patient during treatment planning is needed to ensure patient safety. We offer a freely available collision prediction tool using Blender, a free and open-source 3D computer graphics software toolset. A geometric model of a C-arm linear accelerator including a library of patient models is created inside Blender. Based on the model, collision predictions can be used both to calculate collision-free zones and to check treatment plans for collisions. The tool is validated for two setups, once with and once without a full body phantom with the same tab
SUBMITTER: Guyer G
PROVIDER: S-EPMC10647990 | biostudies-literature | 2023 Nov
REPOSITORIES: biostudies-literature