To develop an angular correction methodology and characterize a high-resolution complementary metal-oxide-semiconductor (CMOS) array for patient specific quality assurance on a robotic arm linear accelerator.Methods
Beam path files from the treatment planning software (TPS) were used to calculate the angle of radiation beam with respect to the detector plane. Beams from multiple discrete angles were delivered to the CMOS detector array and an angular dependency look up table (LUT) was created. The LUT was then used to correct for the angular dependency of the detector. An iso-centric 5 mm fixed cone, non iso-centric multi-target fixed cone, 10 mm Iris and a multi-leaf collimator (MLC) based collimated plan were delivered to the phantom and compared to the TPS with a