ABSTRACT: Background and purpose
: Post-shunt MRI is usually performed at 1.5T under the general assumption that shunt-related susceptibility artifacts would be greater at higher field strengths.Purpose
The purpose is to show that imaging post-shunt idiopathic normal pressure hydrocephalus (iNPH) patients at 3T is feasible and with reduced artifacts as compared to 1.5T.Methods
We manually measured transverse dimensions of artifact at the levels of lateral ventricles, cerebral aqueduct, and cerebellar hemisphere. Areas/volumes of artifacts were calculated assuming an elliptic/ellipsoid shape. Relative extent of shunt-related artifact between field strengths was rated by 3 readers on a 5-point Likert scale. A Wilcoxon Signed Rank Test was used to compare artifact at 1.5T vs 3T