Effect of Robotic-Assisted Gait at Different Levels of Guidance and Body Weight Support on Lower Limb Joint Kinematics and Coordination.
Ontology highlight
ABSTRACT: The Lokomat provides task-oriented therapy for patients with gait disorders. This robotic technology drives the lower limbs in the sagittal plane. However, normative gait also involves motions in the coronal and transverse planes. This study aimed to compare the Lokomat with Treadmill gait through three-dimensional (3D)-joint kinematics and inter-joint coordination. Lower limb kinematics was recorded in 18 healthy participants who walked at 3 km/h on a Treadmill or in a Lokomat with nine combinations of Guidance (30%, 50%, 70%) and bodyweight support (30%, 50%, 70%). Compared to the Treadmill, the Lokomat altered pelvic rotation, decreased pelvis obliquity and hip adduction, and increased ankle rotation. Moreover, the Lokomat resulted in significantly slower velocity at the hip, knee, and