Are all ventilators for NIV performing the same? A bench analysis.


ABSTRACT: Global pandemic due to COVID-19 has increased the interest for ventilators´ use worldwide. New devices have been developed and older ones have undergone a renewed interest, but we lack robust evidence about performance of each ventilator to match appropriate device to a given patient and care environment. The aim of this bench study was to investigate the performance of six devices for noninvasive ventilation, and to compare them in terms of volume delivered, trigger response, pressurization capacity and synchronization in volume assisted controlled and pressure support ventilation. All ventilators were tested under thirty-six experimental conditions by using the lung model ASL5000® (IngMar Medical, Pittsburgh, PA). Two leak levels, two muscle inspiratory efforts and three mechanical patte