Evaluation of a novel CBCT conversion method implemented in a treatment planning system.


ABSTRACT:

Background

To evaluate a novel CBCT conversion algorithm for dose calculation implemented in a research version of a treatment planning system (TPS).

Methods

The algorithm was implemented in a research version of RayStation (v. 11B-DTK, RaySearch, Stockholm, Sweden). CBCTs acquired for each ten head and neck (HN), gynecology (GYN) and lung cancer (LNG) patients were collected and converted using the new algorithm (CBCTc). A bulk density overriding technique implemented in the same version of the TPS was used for comparison (CBCTb). A deformed CT (dCT) was created by using deformable image registration of the planning CT (pCT) to the CBCT to reduce anatomical changes. All treatment plans were recalculated on the pCT, dCT, CBCTc and the CBCT

SUBMITTER: Lechner W 

PROVIDER: S-EPMC10655347 | biostudies-literature | 2023 Nov

REPOSITORIES: biostudies-literature