Accuracy of Crit-Line® monitors in predicting hematocrit and change in blood volume of dogs during hemodialysis.


ABSTRACT:

Background

Ultrafiltration (UF) is an extracorporeal technique for treating fluid overload and is monitored with noninvasive blood volume monitors.

Hypothesis/objectives

The primary objective was to determine the accuracy of Crit-Line® III and IV noninvasive blood volume monitors to estimate canine packed cell volume (PCV) and changes in blood volume during UF. A secondary aim was to determine accuracy of targeted ultrafiltration rates (UFR) compared to calculated delivered UFR.

Methods

An ex vivo study with a single Phoenix® X36 platform and canine packed red blood cells (pRBC) was performed. Through dilution and UF, clinically applicable PCV values were obtained and compared to Crit-Line® hematocrit (Hct). Blood volume was constantly measured and compared to targete

SUBMITTER: Vickery WE 

PROVIDER: S-EPMC10658494 | biostudies-literature | 2023 Nov-Dec