X-ray beam quality after a mirror reflection: Experimental and simulated results for a toroidal mirror in a 4 th generation storage-ring beamline.


ABSTRACT:

Background

The surface errors found in X-ray mirrors constitute a limiting factor for preserving beam quality. This is particularly important when the X-ray beam has low emittance and a significant coherence fraction, like in newly upgraded synchrotron storage rings.

Methods

We studied the fringes observed in the image of an undulator-produced X-ray beam reflected by a high-quality toroidal mirror. The measurements and simulations were performed using different conditions: a photon beam either monochromatic or with large bandwidth, reflected by a mirror with variable curvature.

Results

The experimental data are compared with up-to-date simulation including partial coherence.

Conclusions

The observed fringes in the unfocused beam correlate with low spatial frequency structures in mirror profiles, irrespective of beam coherence. Both classical ray tracing and partially coherent simulations through coherent mode decomposition are confirmed as accurate methods for such simulations.

SUBMITTER: Reyes-Herrera J 

PROVIDER: S-EPMC10660299 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature

altmetric image

Publications

X-ray beam quality after a mirror reflection: Experimental and simulated results for a toroidal mirror in a 4 <sup>th</sup> generation storage-ring beamline.

Reyes-Herrera Juan J   Celestre Rafael R   Cammarata Marco M   Barrett Raymond R   Levantino Matteo M   Sanchez Del Rio Manuel M  

Open research Europe 20231012


<h4>Background</h4>The surface errors found in X-ray mirrors constitute a limiting factor for preserving beam quality. This is particularly important when the X-ray beam has low emittance and a significant coherence fraction, like in newly upgraded synchrotron storage rings.<h4>Methods</h4>We studied the fringes observed in the image of an undulator-produced X-ray beam reflected by a high-quality toroidal mirror. The measurements and simulations were performed using different conditions: a photo  ...[more]

Similar Datasets

| S-EPMC10267196 | biostudies-literature
| S-EPMC5929361 | biostudies-literature
| S-EPMC9588073 | biostudies-literature
| S-EPMC5182027 | biostudies-literature
| S-EPMC6657551 | biostudies-literature
| S-EPMC10234452 | biostudies-literature
| S-EPMC6369583 | biostudies-literature
| S-EPMC8725938 | biostudies-literature
| S-EPMC3362361 | biostudies-literature