ABSTRACT: Background
The surface errors found in X-ray mirrors constitute a limiting factor for preserving beam quality. This is particularly important when the X-ray beam has low emittance and a significant coherence fraction, like in newly upgraded synchrotron storage rings.Methods
We studied the fringes observed in the image of an undulator-produced X-ray beam reflected by a high-quality toroidal mirror. The measurements and simulations were performed using different conditions: a photon beam either monochromatic or with large bandwidth, reflected by a mirror with variable curvature.Results
The experimental data are compared with up-to-date simulation including partial coherence.Conclusions
The observed fringes in the unfocused beam correlate with low spatial frequ