Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring.


ABSTRACT: Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses a motorized press that is used to apply pressure to the fingertip to measure BP. We verified the functionality of the device in a clinical trial (n = 43) resulting in systolic and diastolic pressures ((mean ± SD) mmHg) of (-3.5 ± 8.4) mmHg and (-4.0 ± 4.4) mmHg, respectively. Comparison was made with manual auscultation (n = 26) and automated cuff oscillometry (n = 18). In addition to BP, we demonstrated the ability of the device to assess arterial stiffness (n = 18) and detect atrial fibrillation (n = 6). We were able to introduce a sufficiently small device that could be used for convenient ambulatory measurements with minimal discomfort.

SUBMITTER: Panula T 

PROVIDER: S-EPMC10665806 | biostudies-literature | 2023 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring.

Panula Tuukka T   Sirkiä Jukka-Pekka JP   Koivisto Tero T   Pänkäälä Mikko M   Niiranen Teemu T   Kantola Ilkka I   Kaisti Matti M  

iScience 20231024 11


Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses a motorized press that is used to apply pressure to the fingertip to measure BP. We verified the functionality of the device in a clinical trial (n = 43) resulting in systolic and diastolic pressures  ...[more]

Similar Datasets

| S-EPMC11200005 | biostudies-literature
| S-EPMC8855943 | biostudies-literature
| S-EPMC9856486 | biostudies-literature
| S-EPMC10923543 | biostudies-literature
| S-EPMC6863526 | biostudies-literature
| S-EPMC5703831 | biostudies-literature
| S-EPMC11342199 | biostudies-literature
| S-EPMC6350759 | biostudies-literature
| S-EPMC7074014 | biostudies-literature
| S-EPMC11859408 | biostudies-literature