Ontology highlight
ABSTRACT: Objective
Although needleless connectors (NCs) are widely used in clinical practice, they carry significant risk of bloodstream infection (BSI). In this study, we quantified differences in bacterial transfer and biofilm formation between various NCs.Design
Prospective, clinically simulated in vitro experimental study.Methods
We tested 20 NCs in a 5-day clinical simulation of Staphylococcus aureus inoculations onto NC septum surfaces, which were then flushed with saline and cultured for bacterial transfer. Biofilm formation was measured through destructive sampling of the connector-catheter system. Moreover, 8 NC design factors were evaluated for their influence on bacterial transfer and biofilm formation. This study was designed without a disinfection protoco
SUBMITTER: Ryder M
PROVIDER: S-EPMC10665874 | biostudies-literature | 2023 Nov
REPOSITORIES: biostudies-literature