ABSTRACT: Background
High-power short-duration (HPSD) ablation strategy has emerged as a popular approach for treating atrial fibrillation (AF), with shorter ablation time. The utilized Smart Touch Surround Flow (STSF) catheter, with 56 holes around the electrode, lowers electrode-tissue temperature and thrombus risk. Thus, we conducted this prospective, randomized study to investigate if the HPSD strategy with STSF catheter in AF ablation procedures reduces the silent cerebral embolism (SCE) risk compared to the conventional approach with the Smart Touch (ST) catheter.Methods
From June 2020 to September 2021, 100 AF patients were randomized 1:1 to the HPSD group using the STSF catheter (power set at 50 W) or the conventional group using the ST catheter (power set at 30 to 35 W). Pul