Unknown

Dataset Information

Prosthetic Visual Acuity with the PRIMA System in Patients with Atrophic Age-related Macular Degeneration at 4 years follow-up.


ABSTRACT:

Objective

To assess the efficacy and safety of the PRIMA subretinal neurostimulation system 48-months post-implantation for improving visual acuity (VA) in patients with geographic atrophy (GA) due to age-related macular degeneration (AMD) at 48-months post-implantation.

Design

First-in-human clinical trial of the PRIMA subretinal prosthesis in patients with atrophic AMD, measuring best-corrected ETDRS VA (Clinicaltrials.gov NCT03333954).

Subjects

Five patients with GA, no foveal light perception and VA of logMAR 1.3 to 1.7 in their worse-seeing "study" eye.

Methods

In patients implanted with a subretinal photovoltaic neurostimulation array containing 378 pixels of 100 μm in size, the VA was measured with and without the PRIMA system using ETDRS charts at 1 met

SUBMITTER: Muqit M 

PROVIDER: S-EPMC10680875 | biostudies-literature | 2023 Nov