ABSTRACT: Purpose
To evaluate the accuracy and reproducibility of a patient-specific, customized individually made (CIM) total knee replacement (TKR) using the ORIGIN® prosthesis.Methods
This was a prospective study conducted at a University Hospital from January 15, 2019, to April 30, 2021. The study included patients planned for an ORIGIN® CIM TKR procedure. Exclusion criteria included revision surgery, severe deformity, stiffness, or laxity. Evaluations were carried out using computed tomography scans performed 8 weeks preoperatively and 6 weeks postoperatively. The primary outcome measurements were the preoperative, planned, and postoperative CT scan alignment measurements including the Hip-Knee-Ankle (HKA) angle, mechanical Medial Distal Femoral articular surface Angle (mMDFA, d