Unknown

Dataset Information

Customized-individually-made origin® implants in total knee arthroplasty allow a reliable solution for accurate reproduction of planned implant positioning.


ABSTRACT:

Purpose

To evaluate the accuracy and reproducibility of a patient-specific, customized individually made (CIM) total knee replacement (TKR) using the ORIGIN® prosthesis.

Methods

This was a prospective study conducted at a University Hospital from January 15, 2019, to April 30, 2021. The study included patients planned for an ORIGIN® CIM TKR procedure. Exclusion criteria included revision surgery, severe deformity, stiffness, or laxity. Evaluations were carried out using computed tomography scans performed 8 weeks preoperatively and 6 weeks postoperatively. The primary outcome measurements were the preoperative, planned, and postoperative CT scan alignment measurements including the Hip-Knee-Ankle (HKA) angle, mechanical Medial Distal Femoral articular surface Angle (mMDFA, d

SUBMITTER: Sajan M 

PROVIDER: S-EPMC10684845 | biostudies-literature | 2023 Nov

REPOSITORIES: biostudies-literature